How to
Obtain “Correct” Front-Panel Display for
Agilent 33120A Function/Arbitrary Waveform Generator
The function/arbitrary
waveform generator assumes you are driving a 50-ohm load (this is the power-on
default behavior). If you are driving a high-impedance load (such as the clock
input of a digital gate, or an op-amp input) then you will see that the actual
voltage waveform is two times higher than the displayed value. Follow these
instructions to produce a correct display for high-impedance loads:
Note: Bold face means the
button label, italicized word means button group label
What to do: |
How to do it: |
1. Display the options
menu |
·
Press Shift
(blue button) ·
Press Menu
Þ On/Off |
2. Find and select the
system menu |
·
Rotate the
knob clockwise to select “D: SYS MENU” ·
Press Menu
Þ Ú |
3. Find and select the
output termination submenu |
·
You should
see “1: OUT TERM” ·
Press Menu
Þ Ú |
4. Select load impedance
type |
·
Rotate knob
to select “HIGH Z” |
5. Enter your selection
and return to normal display |
·
Press Menu
Þ Enter |